FAQ for releases before v0.4.0
Note - the information below only applies to releases of the uploader before version 0.4.0
When registering the CNL 2.4 stick, I get a username or password
error, even though I'm sure they're correct
Unfortunately, when an error occurs on the CareLink website, the server does not distinguish between a login error and an internal server error.
Also this error has been seen when a new CNL 2.4 stick is being used, i.e. one that either has not been paired with the pump, or has not been used for an upload to the CareLink website. Check that you have followed all the steps in 600 Series pump and CNL 2.4 preparation
If the problem persists, please make sure you can log into the European CareLink website via a browser on your device before reporting a new issue.
After registering the CNL 2.4 I don't get a username/password error, but there are no Registered Devices
in the list either
This occurs when the CNL 2.4 USB stick has not been detected by the Android device
- Quit the app and unplug and re-plug the CNL 2.4. You should see a pop-up asking to give the app access to the USB (check the "remember" flag)
- Launch the app and press "Get Now". You should get a message that you can't use the USB stick without registering it
- After registering you should see a "Serial Number: xxxxxxx" under "Registered Devices" with a cross next to it
If this still doesn't work, there might be a problem with your USB cable (can you swap it?), or your device may not support USB OTG. Try the USB OTG troubleshooter on the Prerequisites page
Unfortunately there is sometimes misleading information about whether a particular Android device supports USB OTG, and there have been cases where different versions of Android on the same device have given different results.
Please carefully research before updating your Android version as updates cannot usually be reversed.
When I click Get now
the status shows Connecting to Contour Next Link
, then Connected...
but then it says Timeout communicating with Contour Next Link
It's essential that the CNL 2.4 has been used to upload pump data to the CareLink website. This may catch you out if you have been swapping CNL 2.4s.
It's also essential that the current pump was the last pump used to upload data to the CareLink website. This may catch you out if you have received a replacement pump.
Please check that you have followed all the steps in 600 Series pump and CNL 2.4 preparation
-
Please read the Beta Test Programme Requirements in full before proceeding
-
All information, thought, and code described here is intended for informational and educational purposes only. Nightscout currently makes no attempt at HIPAA privacy compliance. Use Nightscout at your own risk, and do not use the information or code to make medical decisions.
-
Use of code from github.com is without warranty or formal support of any kind. Please review this repository's LICENSE for details.
-
All product and company names, trademarks, servicemarks, registered trademarks, and registered servicemarks are the property of their respective holders. Their use is for information purposes and does not imply any affiliation with or endorsement by them.
-
Please note - this project has no association with and is not endorsed by: