Expand Up @@ -88,8 +88,8 @@ In the context of LSSTCam, we extract bright pixels from the dark current, with
Reviewing the differences in bright pixels, we find consistent bright defect counts between Run 6 and Run 7. There appears to be a small excess of bright defects in Run 7.


.. image:: /figures/baselineCharacterization/13557_E1071_BRIGHT_PIXELS_diff.png
:target: /figures/baselineCharacterization/13557_E1071_BRIGHT_PIXELS_diff.png
.. image:: /figures/baselineCharacterization/BRIGHT_PIXELS_13557_E1071_diff.png
:target: /figures/baselineCharacterization/BRIGHT_PIXELS_13557_E1071_diff.png
:alt: The comparison between bright pixel measurements at SLAC and at Cerro Pachon, taking the differences between measurements on a per amp basis.

Taking the difference of defect counts on each amplifier, and separating the amplifiers by the detector manufacturer shows a small excess of bright defects in run 7 when compared to run 6. For ITL sensors, we find 12% of the amplifiers with more bright pixels than run 6. For E2V sensors, we find 4% of the amplifiers with more bright pixels than run 6. Despite this, the number of bright defects between runs does not increase for most sensors.