Hi!
While measuring a cell, it should have done every 50 mV a measurement. Hoewever it seemd to miss some results, although I heard the beep from the measurement (current change).
They are not included in the list or in the csv file. I attached the log file from some measurements.
Same problem here as well. Also, sometimes the discharge just stops with the current setting reset to 0 amps.
This is seemingly caused by the software picking up current fluctuation in Idle mode, and then jumping upon a single unstable current row during the current drop before respectively the current ramp up after internal resistance measurement.