The Patch for v.1.3.3XXX has a problem with calibrations.
When you calibrate the patched app sends the calibrated bg value

image

AND also the non-calibrated value

image

to AAPS within the next update cycle.
This makes AAPS stop closed loop as ther ea re two values coming in at the same time.

The solution would be to compare the timestamp with backfilled FALSE | TRUE.